产品名称:数据记录仪
产品型号:GL7000
产品简介:
新一代数据采集装备
该装备可按需测量需要的信号,并通过添加不同的放大器模块来扩大应用范围。
该装置可附加带触控面板的显示模块,作为独立装置使用或嵌入系统内使用。
1.适用于多种测量的扩展式放达器模块
支持多种测量的放大器模块
该放大器模块支持不同的测量应用,通过添加不同类型的模块,还可以进行混合测量。
Amplifier module
Module |
Voltage |
High Speed Voltage |
Voltage/Temperature |
Logic/Pulse |
Model number |
GL7-V |
GL7-HSV |
GL7-M |
GL7-L/P |
Number of input channels |
10 channels |
4 channels |
10 channels |
16 channels |
Input terminal |
Screw terminal (M3) |
BNC connector |
Screw terminal (M3) |
Specific connector |
Input method |
All channel isolated unbalanced input, simultaneous sampling |
All channels isolated balanced input, scanning channels for sampling |
All channels common ground, simultaneous sampling |
Sampling speed |
1 k Samples/s to 1 Sample/h (1ms to 1hr. interval) |
1 M Samples/s to 1 Sample/h (1μs to 1hr. interval) |
100 Samples/s at 1-10ch to 1 Sample/h (10ms at 1-10ch to 1hr. interval) |
Logic mode: up to 1 M Samples/s (1μs interval) Pulse mode: up to 10 k Samples/s (100μs interval) |
Measurement range |
100mV to 100V F.S., and 1-5V F.S. |
Volt.: 20mV to 50V F.S., and 1-5V F.S. Temp.:Thermocouple: K, J, E, T, R, S, B, N, W (WRe5-26) RTD: Pt100 (IEC751), Pt1000 (IEC751), Pt100J (JIS) |
Bi-level signal, up to 24V Select either Logic or Pulse mode Logic: Signal pattern Pulse: Counting in Instant, Accumulating, Rotation (max. 15 million count) |
A/D converter |
Successive Approximation, 16 bits |
Sigma-delta, 16 bits |
|
Withstand voltage |
Between channels: 1000 V, 1min. Between inputs and GND: 1000 V, 1 min. |
Between channels: 350 V, 1 min. Between inputs and GND: 350 V, 1 min. |
|
Built-in RAM |
2 million samples |
即使在放大器模块增加的情况下,也可保持采样速度
即使增加模块数量,电压*与电压/温度放大器可在不影响采样速度的情况下,保持高速多通道测量。
Sampling speed
Amplifier Module |
Channels in 1 module |
Max. sampling speed in the module |
Media type to save data |
Max. sampling speed in the GL7000 |
Attached to 1 or 2 modules |
Attached to3 or 4 modules |
Attached to 5 to 10 modules |
Voltage Module (GL7-V) |
10 channels |
1 k Samples/s (1ms interval) |
Built-in RAM |
1 k Samples/s (1ms interval) |
Built-in Flash |
SD card |
SSD *3 |
Volt./Temp. Module (GL7-M) |
10 channels |
100 Samples/s (10ms interval) |
Built-in RAM |
100 Samples/s (10ms interval) |
Built-in Flash |
SD card |
SSD *3 |
High-speed voltage Module (GL7-HSV) |
4 channels |
1 M Samples/s (1μs interval) |
Built-in RAM |
1 M Samples/s (1μs interval) |
Built-in Flash |
1 k Samples/s (1ms interval) |
SD card |
SSD *3 |
1 M S/s (1μs interval) |
500 k S/s (2μs interval) |
200 k S/s (5μs interval) |
Logic/Pulse Module (GL-L/P) |
16 channels |
In Logic mode,1 M Samples/s (1μs interval) |
Built-in RAM |
1 M Samples/s (1μs interval) *1 |
Built-in Flash |
1 k Samples/s (1ms interval) *1 |
SD card |
SSD *3 |
1 M S/s (1μs interval) |
500 k S/s (2μs interval) |
200k S/s (5μs interval) *1 |
In Pulse mode,10 k Samples/s (100μs interval) |
Built-in RAM |
1 M S/s (1μs interval) |
Not Available *2 |
Built-in Flash |
1 k S/s (1ms interval) |
SD card |
SSD *3 |
1 M S/s (1μs interval) |
放大器最多可附加10个模块
可附加多达10个放大器模块,进行多通道测量,一台GL7000的通道上限为112.
2.配触控面板的高清显示模块,可实现独立操作或系统嵌入式操作
可实现独立式配置或系统嵌入式配置的拆卸式显示模块
在不适用PC的情况下,通过附加的显示模块即可完成测量设置与信号测量。显示模块可安装在不同的位置,使用一条LAM线缆*与主机连接,进行远程操作;该模块还可嵌入系统内部。即使在连接显示模块的情况下,主机依然可由PC操作。
高清显示与触控面板提高使用的便捷性
触控面板使条件设置更直观,还可使用类似GL系列的游标键进行操作。
5.7英寸高清易读的大屏LCD显示屏
使用清晰的5.7英寸彩色宽屏TFT-LCD显示屏(VGA:640*480点阵)。无论是泼形模式还是数字模式,均可轻松读取数据,检查测量参数设置。
3.技术参数
Voltage/Temperature Module Specifications
Model number |
GL7-M |
Number of input channels |
10 channels |
Input method |
All channels isolated balanced input, Scans channels for sampling, Screw terminal (M3 screw) |
Sampling speed |
100 Samples/s at 1-10ch to 1 Sample/h (10 ms with 1-10ch to 1 hr. interval) |
Measurement range |
Voltage |
20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50 V, and 1-5 V/F.S. |
Temperature |
Thermocouple: K, J, E, T, R, S, B, N, and W (WRe5-26), RTD: Pt100, JPt100 (JIS), Pt1000 (IEC751) |
Humidity *1 |
0 to 100 % (using scaling function in 5V range, humidity sensor B-530) |
Measurement accuracy*2 |
Voltage |
± 0.1 % of F.S. |
Temperature |
Thermocouple |
Measurement range |
Measurement accuracy |
R/S |
0 °C ≤ TS ≤ 100 °C |
± 5.2 °C |
100 °C < TS ≤ 300 °C |
± 3.0 °C |
R: 300 °C < TS ≤ 1600 °C |
± (0.05 % of reading + 2.0 °C) |
S: 300 °C < TS ≤ 1760 °C |
± (0.05 % of reading + 2.0 °C) |
B |
400 °C ≤ TS ≤ 600 °C |
± 3.5 °C |
600 °C < TS ≤ 1820 °C |
± (0.05 % of reading + 2.0 °C) |
K |
-200 °C ≤ TS ≤ -100 °C |
± (0.05 % of reading + 2.0 °C) |
-100 °C < TS ≤ 1370 °C |
± (0.05 % of reading + 1.0 °C) |
E |
-200 °C ≤ TS ≤ -100 °C |
± (0.05 % of reading + 2.0 °C) |
-100 °C < TS ≤ 800 °C |
± (0.05 % of reading + 1.0 °C) |
T |
-200 °C ≤ TS ≤ -100 °C |
± (0.1 % of reading + 1.5 °C) |
-100 °C < TS ≤ 400 °C |
± (0.1 % of reading + 0.5 °C) |
J |
-200 °C ≤ TS ≤ -100 °C |
± 2.7 °C |
-100 °C < TS ≤ 100 °C |
± 1.7 °C |
100 °C < TS ≤ 1100 °C |
± (0.05 % of reading + 1.0 °C) |
N |
0 °C ≤ TS ≤ 1300 °C |
± (0.1 % of reading + 1.0 °C) |
W |
0 °C ≤ TS ≤ 2000 °C |
± (0.1 % of reading + 1.5 °C) |
Reference Junction Compensation (R.J.C.) accuracy: ± 0.5 °C *3 |
RTD |
Measurement range |
Driving current |
Accuracy |
Pt100 |
-200 °C to 850 °C (FS = 1050 °C) |
1 mA |
± 1.0 °C |
JPt100 |
-200 °C to 500 °C (FS = 700 °C) |
1 mA |
± 0.8 °C |
Pt1000 |
-200 °C to 500 °C (FS = 700 °C) |
0.2 mA |
± 0.8 °C |
R.J. Compensation |
Selecting of the internal or external |
A/D Converter |
Sigma-Delta type, 16 bits (effective resolution: 1/40000 of measuring full range) |
Stability with temperature |
Gain |
0.01 % of F.S./°C |
Zero *4 |
0.02 % of F.S./°C |
Input impedance |
1 MΩ ± 5 % |
Maximum input voltage |
Between (+) / (-) terminal |
60 V p-p |
Between channels |
60 V p-p |
Between channel / GND |
60 V p-p |
Maximum voltage |
Between channels |
350 V p-p (1 minute) |
Between channel / GND |
350 V p-p (1 minute) |
Isolation |
Between input / GND |
Min. 50 MΩ (at 500 V DC) |
Common-mode rejection ratio |
Min. 90 dB (50/60 Hz, Signal source impedance: Max. 300 Ω) |
Filter |
Off, 2, 5, 10, 20, 40 (Moving average in selected number. When the sample is longer than 5 seconds, the data sampled in the sub-sample (5 seconds) will be used for creating the average value. ) |
5V output |
Driving the humidity sensor B-530, 1 channel |
External dimensions (W×D×H) |
49 x 136 x 160 mm (Excluding projection) |
Weight |
Approx. 770 g |
Voltage module specifications
Voltage module specifications |
Voltage |
High Speed Voltage |
Model number |
GL7-V |
GL7-HSV |
Number of input channels |
10 channels |
4 channels |
Input method |
All channels isolated unbalanced input, Simultaneous sampling, Screw terminal |
All channels isolated unbalanced input, Simultaneous sampling, BNC connector |
Sampling speed (interval) |
1 k Samples/s to 1 Sample/h (1ms to 1h) |
1 M Samples/s to 1 Sample/h (1μs to 1h) |
Measurement range |
100, 200, 500 mV,1, 2, 5, 10, 20, 50, 100 V, and 1-5 V/F.S. |
Measurement accuracy*5 |
± 0.25 % of F.S. |
A/D Converter |
Successive Approximation type, 16 bits (effective resolution: 1/40000 of measuring full range) |
Stability with temperature |
Gain |
0.01 % of F.S./°C |
Zero |
0.02 % of F.S./°C |
Input impedance |
1 MΩ ± 5 % |
Maximum input voltage |
Between (+) / (-) terminal |
100mV to 1V range: 60 V p-p, 2V to 100V range: 100 V p-p |
Between channels |
60 V p-p |
Between channel / GND |
60 V p-p |
Maximum voltage |
Between channels |
1000 V p-p (1 minute) |
Between channel / GND |
1000 V p-p (1 minute) |
Isolation |
Between input / GND |
Min. 50 MΩ (at 500 V DC) |
Common-mode rejection ratio |
Min. 90 dB (50/60 Hz, Signal source impedance: Max. 300 Ω) |
Frequency response |
DC to 1 k Hz (at +1/-3 dB) |
DC to 2 k Hz (at +1/-3 dB) |
Filter (Low pass) |
Off, Line(1.5Hz), 5, 50, 500 Hz (-3dB, 6dB/oct) |
Off, Line(1.5Hz), 5, 50, 500, 5k, 50k Hz (at -3dB, 6dB/oct) |
External dimensions (W×D×H) |
49 x 136 x 160 mm (Excluding projection) |
Weight |
Approx. 840 g |
Approx. 740 g |
Logic/Pulse Module Specifications
Model number |
GL7-L/P |
Number of input channels |
16 channels |
Input method |
All channels common ground, simultaneous sampling, Circular connector (4ch/connector) |
Sampling speed |
Logic mode |
Up to 1 M Samples/s (1μs interval) |
Pulse mode |
Up to 10 k Samples/s (100μs interval) |
Measurement mode |
Selecting of the Logic input mode or Pulse input mode *6 |
Mode |
Pulse |
Rotation count (RPM), Accumulating count, Instant count |
Rotation count (RPM) |
Function |
Counting the number of pulses per sampling interval and then it is converted to RPM |
Range |
50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M rpm/F.S. |
Accumulating count |
Function |
Accumulating the number of pulses from the start of measurement |
Range |
50, 500, 5 k, 50 k, 500 k, 5 M, 50 M, 500 M counts/F.S. |
Instant count |
Function |
Counting the number of pulses per sampling interval (count is reset at each sampling) |
Range |
50, 500, 5 k, 50 k, 500 k, 5 M, 50 M, 500 M counts/F.S. |
Max. input frequency |
1 M Hz |
Max. number of count |
15 M counts (24 bits counter is used) |
Input signal |
Voltage range |
0 to +24 V (common ground) |
Signal type |
Contact (Relay), Open collector, Voltage |
Threshold |
Approx. 2.5 V |
Hysteresis |
Approx. 0.5 V (2.5 V to 3 V) |
Filter |
Off or On (-3 dB at 50 Hz) |
External dimensions (W×D×H) |
49 x 136 x 160 mm (Excluding projection) |
Weight |
Approx. 700 g |
|